Product description
22 parameters with P_LCC, P_LCR and RBC, WBC, PLT histograms
Micro sample volume at 13μL for whole blood mode and 20μL for pre-dilute mode
2 reagent system (diluent & lyse), internal storage of lyse
On board storage capacity of 100,000 samples results
RFID reagent close system optional
Support USB data backup & system upgrade
Build-in thermal printer
LIS and HIS available with HL7 protocol
Small in size and light in weight
Specification
| Principles | WBC, RBC and PLT: Impedance |
| HGB:photoelectric colorimetry |
| Parameters | 22 reportable parameters with P-LCC, P-LCR and RBC, WBC, PLT histograms |
| 3 histograms for WBC, RBC and PLT |
| Reagent | Diluent, lyse |
| Sample Volume | Prediluted mode 20 uL |
| Whole blood mode 13 uL |
| Capillary whole blood mode 13 uL |
| Throughput | 72 samples per hour with double channel design |
| Diluention Ratio |
| Whole Blood | Capillary Blood |
| WBC/HGB | 1:232 | 1:400 |
| RBC/PLT | 1:40000 | 1:45000 |
| QC Mode | XJ, X, X-R, X-B |
| Display | 10.4 inch Touch Screen |
| Data Storage Capacity | Up to 100,000 results including numeric and graphical information |
| Communication | LAN Port supports HL7 protocol |
| Interface | 6 USB(mouse, keyboard, HP printer, barcode reader and flash disk), LAN, Support bi-directional LIS |
| Operating Environment | Temperature: 10°C~30°C; Humidity: 20%~85%; Air pressure: 70kPa~106kPa; |