| Main body | High-strength engineering plastic rack, light, beautiful and corrosion resistant
 | 
| Scope of application | adult, child, veterinary | 
| Gas source | O2: 0.27~0.55MPa | 
| Flow meter | O2: 0.1~2L/min    2.5~10L/min | 
| Flow rate of rapid oxygen supply | 25~75L/min | 
| Low oxygen pressure alarm | There will be sound alarm when the oxygen pressure < 0.2MPa
 | 
| Vaporizer | It has the function of automatic compensation based on pressure,temperature,and flow rate.The regulation range of evaporator concentration is 0~5 vol%.Among Halothane,Enflurane,Isoflurane and Sevoflurane,one can be chosen for application as required by the customer. | 
| Respiratory circuit | working mode: all-close, semi-close, semi-open releasing pressure: 0.5~7 kPa
 | 
| Respiratory bellows | bellows for adults, bellows for children tidal volume range:0~1500 ml
 | 
| Ventilator | 
| Display mode |  | 
| Ventilation mode | IPPV, MANUAL (Manual monitoring of tidal volume, ventilation volume, respiratory rate)
 | 
| Ventilation parameters | Tidal volume                50~1500ml Rate                       2~99 bpm
 SIMV rate                  2~20 bpm
 I:E                        2:1~1:8
 Inspiratory trigger pressure     -1.0~2.0 kPa
 Pressure range               1.0~6.0 kPa
 | 
| Parameters for ventilation monitoring | tidal volume, ventilation volume, IPPV rate, I/E, peak pressure of airway, inspiratory
 trigger pressure, pressure - time waveform
 | 
| Security Alarm System | 
| Airway pressure alarm | upper limit setting range        1.0~6.0 kPa low limit setting range          0.4~2.0 kPa
 | 
| Per-minute ventilation volume alarm | upper limit setting range        3.0~30L/min low limit setting range          1.0~10L/min
 | 
| Sustained high-pressure alarm | it will give alarm when stress have consistently been higher than 2.5 kPa
 | 
| Suffocation alarm | it will give sound and light alarm if there is no tidal volume input in 15s
 | 
| Power alarm Power
 | AC 220V±10%     50Hz±1Hz.
 |