Main body | High-strength engineering plastic rack, light, beautiful and corrosion resistant |
Scope of application | adult, child |
Gas source | O2: 0.27~0.55MPa N2O: 0.27~0.55MPa |
Flow meter | O2: 0.1~10L/min N2O: 0.1~10L/min |
O2, N2O linkage and N2O Stopper When using nitrous oxide, oxygen concentration>25%; When the oxygen pressure<0.2MPa, the flow of nitrous oxide would be cut off. |
Flow rate of rapid oxygen supply | 25~75L/min |
Low oxygen pressure alarm | There will be sound alarm when the oxygen pressure < 0.2MPa |
Vaporizer | It has the function of automatic compensation based on pressure, temperature, and flow rate. The regulation range of evaporator concentration is 0~5 vol%. Among Halothane, Enflurane, Isoflurane and Sevoflurane, one can be chosen for application as required by the customer. And vaporizers that are imported with original packaging are also available. |
Respiratory circuit | working mode: all-close, semi-close, semi-open releasing pressure: 0.5~7 kPa |
Respiratory bellows | bellows for adults, bellows for children tidal volume range:0~1,500 ml |
Ventilator |
Display mode | High-definition 5.7” LCD screen display |
Ventilation mode | IPPV, SIPPV, SIMV, PEE,MANUAL, SIGH |
Ventilation parameters | Tidal volume | 50~1500ml |
Rate | 2~99bmp |
SIMV rate | 2~20bpm |
I:E | 2:1~1:8 |
Inspiratory trigger pressure | -1.0~2.0 kPa |
PEEP | 0~2.0 kPa |
Pressure range | 1.0~6.0 kPa |
SIGH | 1.5 times the inspiratory time |
Parameters for ventilation monitoring tidal volume, ventilation volume, IPPV rate, SIMV rate, total respiratory rate, I/E, peak pressure of airway, pressure - time waveform, flow rate - time waveform, PEEP, inspiratory trigger pressure |
Monitoring of oxygen concentration | (21%~100%) |
Security Alarm System |
Oxygen Concentration alarm | upper limit setting range (50%~100%) |
low limit setting range (15%~50%) |
Airway pressure alarm | upper limit setting range 1.0~6.0 kPa |
low limit setting range 0.4~2.0 kPa |
Per-minute ventilation volume alarm | upper limit setting range 3.0~30L/min |
low limit setting range 1.0~10L/min |
Sustained high-pressure alarm | it will give alarm when stress have consistently been higher than 2.5 kPa |
Suffocation alarm | it will give sound and light alarm if there is no tidal volume input for 15 seconds |
Power alarm |
Power | AC 220V±10% 50Hz±1Hz (UPS, Storage battery) |
Items within brackets are optional. |