| Main body | High-strength engineering plastic rack, light, beautiful and corrosion resistant | 
| Scope of application | adult, child | 
| Gas source | O2: 0.27~0.55MPa   N2O: 0.27~0.55MPa | 
| Flow meter | O2: 0.1~1.0L/min   1.1~10L/min | 
| N20: 0.1~1.0L/min   1.1~10L/min | 
| O2, N2O linkage and N2O Stopper When using nitrous oxide, oxygen concentration>25%;
 When the oxygen pressure<0.2Kpa, the flow of nitrous oxide would be cut off.
 | 
| Flow rate of rapid oxygen supply | 25~75L/min | 
| Low oxygen pressure alarm 
 | There will be sound alarm when the oxygen pressure < 0.2MPa
 | 
| Vaporizer | it has the function of automatic compensation based on pressure, temperature, and flow rate. The regulation
 range of evaporator concentration is 0~5 vol%. Among
 Halothane,Enflurane, Isoflurane and Sevoflurane, one
 can be chosen for application as required by the
 customer. And vaporizers that are imported with original
 packaging are also available.
 | 
| Respiratory circuit 
 | working mode: all-close, semi-close, semi-open releasing pressure: 0.5~7 kPa
 | 
| Respiratory Bellows 
 | bellows for adults, bellows for children tidal volume range:0~1,500 ml
 | 
| Ventilator | 
| Display mode | High-definition TFT color LCD screen display | 
| Ventilation mode 
 | IPPV, SIPPV, SIMV, PEEP, MANUAL, SIGH, (IMV, VCV, PCV)
 | 
| Ventilation parameters | Tidal volume | 50~1500 ml (20~1500 ml) | 
| Rate | 1~99 bpm | 
| SIMV rate | 2~20 bpm | 
| I:E | 2:1~1:8 (4:1~1:8) | 
| Inspiratory trigger pressure | -1.0~2.0Kpa | 
| PEEP PressureRange
 | 0~2.0Kpa 1.0~6.0Kpa
 | 
| SIGH | 1.5 times the inspiratory time (70-150 adjustable) | 
| Inspiratory platform | 0~50% | 
| Parameters for ventilation monitoring tidal volume, ventilation volume, IPPV rate, SIMV rate, total respiratory rate, I/E, peak pressure of airway, average pressure, pressure - time waveform, flow rate - time waveform, PEEP, inspiratory trigger pressure, Inspiratory platform (autonomous inspiratory rate, lung compliance, airway resistance,flow-volume oop,pressure-volume loop) | 
| Monitoring of oxygen concentration | 21%~100% | 
| Security Alarm System | 
| Oxygen Concentration alarm | upper limit setting range 21%~100% | 
| low limit setting range   10%~80% | 
| Airway pressure alarm | upper limit setting range 0.3~6.0 kPa | 
| low limit setting range   0.2~5.0 kPa | 
| Per-minute ventilation volume alarm | upper limit setting range 3.0~30L/min | 
| low limit setting range   1.0~10L/min | 
| Sustained high-pressure alarm | it will give alarm when stress have consistently been higher than 2.5 kPa | 
| Suffocation alarm | it will give sound and light alarm if there is no tidal volume input for 15 seconds
 | 
| Power alarm | 
| Power | AC 220V±10%     50Hz±1Hz(UPS, Storage battery) | 
| (Mechanical Arms)(End-expiratory CO2 monitoring) | 
| Items within brackets are optional. |