100mA Medical X-ray equipment CCX-100
USE
General radiography, Filter-ray-filter radiography.
Unit Configuration
Control, HT generator unit, Ht cable, X-ray tube unit, Radiographic table-X-ray source column, Collimator.
Function
This photography x ray can used for general radiography, Bucky radiography of body's head, chest, foot and hand, bones in lying.
Feature of x-ray machine:
1. meter-indicating.
2. single-table single tube.
3. single-phase, full wave rectification
4. stationary anode x-ray tube assembly.
5. X-ray tube filament s voltage regulator, space charge compensator .
6. Radiograph plain table wimple structure and easy assembling.
Min Technical Indexes
Item | Contents | Index |
Demand of power supply | Capacity | Two-phase 30KVA |
Voltage, Frequency | 380V/220V ±10%50HZ ±0.5Hz |
Intemal resistance | 2Ω(380V) 0.6 Ω(220V) |
power supply | Adjustable range | 380±10% at220V being adjustable continuously |
Radiograph | Voltage | 50-90KV |
Current | Small focus: 15mA,Big focus:30mA 60mA 100mA |
Time | 0.04~6.35s 23grades together grading a according to R10 coefficient |
High-voltage generator | Capacity | 7KVA |
Max DC output voltage | 90KV |
Max DC output current | 100mA |
X-Ray tube | Model | 0XD4-2.9/100 |
Focus | Small focus 1.8×1.8mm big focus4.3×4.3mm |
Radiographic table | Length×width×height of bed surface | 2000mm×665mm×700mm |
Moving range of bed surface | 640mm in Length,200mm in breadth |
Ray-filter of table | Travel in length ≥50mm |
Grid density N=28 |
Convergence distance f.=100cm |
Grid ratio r=8 |
Column of X-ray tube unit | Moving Longitudinally along | 1000mm or 1800mm |
Radiographic table |
Distance to friend as moving up and down (along column) | 650~1800mm |
Rotary around the center of cross arm | ±180° |
Rotary around the shaft line of X-ray annular tubes | -10~60~+120° |
Max size of cassette | 356mm×432mm(14″×17″) |