| Main body | High-strength engineering plastic rack, light, beautiful and corrosion resistant |
| Scope of application | adult, child |
| Gas source | O2: 0.27~0.55MPa N2O: 0.27~0.55MPa |
| Flow meter | O2: 0.1~10L/min N2O: 0.1~10L/min |
O2, N2O linkage and N2O Stopper When using nitrous oxide, oxygen concentration>25%; When the oxygen pressure<0.2MPa, the flow of nitrous oxide would be cut off. |
| Flow rate of rapid oxygen supply | 25~75L/min |
| Low oxygen pressure alarm | There will be sound alarm when the oxygen pressure < 0.2MPa |
| Vaporizer | It has the function of automatic compensation based on pressure, temperature, and flow rate. The regulation range of evaporator concentration is 0~5 vol%. Among Halothane, Enflurane, Isoflurane and Sevoflurane, one can be chosen for application as required by the customer. And vaporizers that are imported with original packaging are also available. |
| Respiratory circuit | working mode: all-close, semi-close, semi-open releasing pressure: 0.5~7 kPa |
| Respiratory bellows | bellows for adults, bellows for children tidal volume range:0~1,500 ml |
| Ventilator |
| Display mode | High-definition 5.7” LCD screen display |
| Ventilation mode | IPPV, SIPPV, SIMV, PEE,MANUAL, SIGH |
| Ventilation parameters | Tidal volume | 50~1500ml |
| Rate | 2~99bmp |
| SIMV rate | 2~20bpm |
| I:E | 2:1~1:8 |
| Inspiratory trigger pressure | -1.0~2.0 kPa |
| PEEP | 0~2.0 kPa |
| Pressure range | 1.0~6.0 kPa |
| SIGH | 1.5 times the inspiratory time |
| Parameters for ventilation monitoring tidal volume, ventilation volume, IPPV rate, SIMV rate, total respiratory rate, I/E, peak pressure of airway, pressure - time waveform, flow rate - time waveform, PEEP, inspiratory trigger pressure |
| Monitoring of oxygen concentration | (21%~100%) |
| Security Alarm System |
| Oxygen Concentration alarm | upper limit setting range (50%~100%) |
| low limit setting range (15%~50%) |
| Airway pressure alarm | upper limit setting range 1.0~6.0 kPa |
| low limit setting range 0.4~2.0 kPa |
| Per-minute ventilation volume alarm | upper limit setting range 3.0~30L/min |
| low limit setting range 1.0~10L/min |
| Sustained high-pressure alarm | it will give alarm when stress have consistently been higher than 2.5 kPa |
| Suffocation alarm | it will give sound and light alarm if there is no tidal volume input for 15 seconds |
| Power alarm |
| Power | AC 220V±10% 50Hz±1Hz (UPS, Storage battery) |
| Items within brackets are optional. |